Standards
GB/T 3536, ASTM D92
Application
The instrument is used to determine the flash points of the petroleum products with flash points above 79℃ and below 400℃ except fuel oils. It is a high-end automatic instrument.
Standards
GB/T 3536, ASTM D92
Application
The instrument is used to determine the flash points of the petroleum products with flash points above 79℃ and below 400℃ except fuel oils. It is a high-end automatic instrument.
Main technical specifications
1. Power supply: AC 220V±10%, 50Hz/60Hz
2. Maximum power consumption: 600W
3. Measuring range: 75 - 410℃
4. Temperature resolution: 0.1℃
5. Repeatability&reproducibility: ≤6℃(R) & 12℃(r)
6. Flash point detecting unit: Ionization ring detection
7. Temperature sensor: Pt100, Imported
8. Heating rate: 14-17℃/min & 5-6℃/min
9. Ignition method: Gas or electronic, switchable
10. Gas source: LPG, butane gas, natural gas (pressure <10 kPa)
11. Display: 7-inch color touch LCD
12. Data storage: Over 10000pcs
13. Data output: USB/LIMS(customizable)
14. Communication port: RS485, RJ45
15. Safety functions: Automatic shutdown and alarm upon reaching the limit temperature/Forced air cooling/Gas spraying extinguishment
16. Working condition: 10-35℃, RH≤85%
17. Dimension and net weight: 265 mm × 525 mm × 365 mm, 15Kg
Main technical features
1. 7 inch color touch LCD. Built-in computer control. English display. It can show all test parameters on the screen. Easy to use.
2. The heating, test arm lifting-up, test arm lowering-down, flash point measuring and forced air cooling are all automatically finished. After the test procedure is over, the result can be printed by the built-in micro printer directly.
3. Built-in BARO sensor. It can correct the test result according to the detected atmospheric pressure automatically.
4. It can save at least 10000pcs of test results. RS-485,RJ45 ports are equipped on the instrument. The test data can also be exported by an USD disk. Further more, the test data can be uploaded to LIMS(Model SDB-3536Z-Pro, customizable).
5. Gas ignition and electric ignition are switchable.
6. Kinds of safety functions are set up in the instrument for a safer operation to the user.
Petroleum Testing Instruments
Petroleum Testing Instruments
SDB-261Z Automatic Pensky-Martens Closed Cup Flash Point Tester
Petroleum Testing Instruments
Petroleum Testing Instruments
Petroleum Testing Instruments
SDB-6536C Low Temperature Lab Distillation Apparatus (Double Unit)
Petroleum Testing Instruments
Petroleum Testing Instruments
Petroleum Testing Instruments