Standards
GB/T 261, ASTM D93
Application
This instrument is used to determine the flash point of petroleum products in the temperature range from 40℃ to 300℃.
Standards
GB/T 261, ASTM D93
Application
This instrument is used to determine the flash point of petroleum products in the temperature range from 40℃ to 300℃.
Main technical specifications
1. Power supply: AC(220±10%)V, 50Hz.
2. Heating device:The heating power is continuously adjustable from 0 W to 600W.
3. Heating rate: (1~12)℃/min;Controllable and adjustable.
4. Stirring rate: Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
5. Oil cup:(1) Inner diameter: 50.7mm~50.8mm.
(2) Depth: 55.7mm~56.0mm
(3) The scribed line depth of capacity of testing oil: 33.9mm~34.3mm.
(4) Capacity of testing oil:about 70ml
6. Igniting device:(1) Igniting source: gas (or other civilian fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Thermometers: Mercury-in-glass thermometer.Specifications are as below:
(1) Scale -5℃~110℃, division 0.5℃。
(2) Scale 20℃~150℃, division 1℃。
(3) Scale 90℃~370℃, division 2℃。
8. Ambient temperature:≤ 35℃
9. Relative humidity:≤ 85%
10.Maximum power consumption:650W
Main technical features
1. Digital displays the voltage value. Easy to control the heating rate.
2. Classical design. The performance is stable.
Petroleum Testing Instruments
Petroleum Testing Instruments
Petroleum Testing Instruments