Description

Standards

SH/T 0193, ASTM D2272

Application

The instrument is used to determine the oxidation stability of steam turbine with the same composition (base oil and additive) . It can also be used to determine new mineral insulating oil containing 2, 6-BHT.

Main technical specifications

1. Power supply: AC220V±10%
2. Heating power: 2500W
3. Pressure sensor: 0~1.6MPa, precision ±2‰
4. Temperature range: Ambient~200.0℃, adjustable.
5. Temperature precision: ±0.1℃
6. Motor: 100±5r/min
7. Included angle between oxygen bomb and water level: 30°

Main technical features

1.The metal block bath is used to decrease the smelly gas during the test. No need to use liquid medium.
2.Single bomb is tested independently. 2 instruments can be operated separately at most at the same time. The effiency is highly improved.
3.The ventilation is easy. No need to put the instrument into the fume cupboard. The exhaust gas can be moved out by a plastic pipe.
4. The temperature of sample can be tested by the temperature sensor during the test.