SDB-3536A Automatic Cleveland Open Cup Flash Point Tester

  • Description
  • Specifications&Features
  • Inquiry

Standards

GB/T 3536, ASTM D92

Application

The instrument is used to determine the flash points of the petroleum products with flash points above 79℃ and below 400℃ except fuel oils.

Main technical specifications

1. Power supply:AC 220V±10%, 50 HZ
2. Maximum power consumption:500W
3. Measurement range:Ambient~350℃
4. Temperature precision:±0.1℃
5. Repeatability&reproducibility:In accordance with GB/T 3536 and ASTM D92
6. Heat source:Civil gas
7. Heating rate:In accordance with GB/T 3536 and ASTM D92
8. Data saving:120 pcs of historical records
9. Printing: Built-in Printer
10.Cooling down:Forced air cooling
11.Ambient temperature:10℃~35℃
12.Relative humidity:≤85%

Main technical features

1. Colored touch-screen operation. Prompt menu and curve are shown on the screen.
2. The heating, test arm lifting-up, test arm lowering-down, flash point detecting and air cooling are all automatic.
3. This instrument can avoid the influence of atmospheric pressure and calculate the correction value automatically.