Standards
SH/T 0059, ASTM D5800 procedure A

Application

The instrument is used to determine the evaporation loss of lubricating oils(particularly engine oils).

Main technical specifications

1. Power supply: AC220V±10%, 50Hz
2. Operation mode: Colored touch screen operation.PID temperature control.
3. Temperature control range: Ambient~250℃
4. Temperature control precision:±0.5℃
5. Heating mode: Wood-metal bath
6. Pressure adjustment: High precision needle valve
7. Filtering mode: Vacuum pump
8. Differential manometer: International brand (inclined manometer is optional)
9. Maximum power consumption: 1.6Kw
10.Overall dimensions: 350mm×280mm×420mm (main unit)

Main technical features 

1. Wood-metal bath heating unit provides good heating performance.
2. International brand differential manometer ensures the correct pressure.
3. Vacuum filtering system is equipped.
4. A digital timer is built in for recording the test time.
5. The evaporation crucible can meet requirement of ASTM D5800
6. International brand needle valve and precise pressure regulation system makes the flow correct.
7. Colored touch LCD operation.Easy to use.